SEDECAL Mobile X-Ray Unit
- Motorized with Smooth Control
- High Frequency X-Ray Generator
- 125KVP (20 kW) or 150 KVP (30 kW) available
- 0.001 sec to 8 sec Exposure Time
- Easy to position, Triple Articulated and Counter Balanced X-Ray Tube
- Requires Standard 240Vac/10amp power
|

|
Specifications for High Frequency X-Ray Generator
| Constant Potential kW Ratings |
20 (SM-20HF) |
30 (SM-20HF) |
| kVp Range in 1 kVp Steps |
40 to 125 |
40 to 150 |
| Accuracy |
± (5% + 1kVp) |
± (5% + 1kVp) |
| mA Stations |
10, 12'5, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320 |
10, 12'5, 16, 20, 25, 32, 40, 50, 64, 80, 100, 125, 160, 200, 250, 320, 400, 500 |
| Accuracy |
± (5% + 0.1mA) |
± (5% + 0.1mA) |
| Exposure Time Range |
1 ms to 8 seconds |
1 ms to 8 seconds |
| Accuracy |
± (1% + 0.1ms) |
± (1% + 0.1ms) |
| mAs Range |
0.1 mAs to 500 mAs |
0.1 mAs to 500 mAs |
| Battery mAs Capacity |
52.500 mAs |
52.500 mAs |
| Automatic Collimation |
Standard |
Standard |
| Anatomical Programming |
Standard |
Standard |
| Operation Independent from Mais Supply |
Standard |
Standard |
Automatic mAs Compensation
Depending Focal Distance (Radiation) |
Standard |
Standard |
| Automatic Exposure Control |
Optional |
Optional |
| Connection for Bucky |
Standard |
Standard |
| Remote Hand-switch |
Optional |
Optional |
| Line Voltage Range and Phases |
110-240 VAC Single phase |
110-240 VAC Single phase |
| X-Ray Tubes (Rotating Anode) |
|
|
| Focal Spots |
0,6 / 1,2 |
0,6 / 1,2 |
| KHU/kVp |
140 / 150 |
140 / 150 |
| Target Angle |
12° |
12° |
| Optional Tube |
|
|
| Focal Spots |
|
0,6 / 1,2 |
| KHU/kVp |
|
300 / 150 |
| Target Angle |
|
12° |
| Weight |
470 kg |
470 kg |
|
Specifications subject to change without notice
|
|